Model Number 1009-9050-000 |
Device Problem
Computer Software Problem (1112)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Customer contact reports no patient information is available.Unique identifier: (b)(4).Ge healthcare's investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.
|
|
Event Description
|
The hospital reported a malfunction causing a loss of mechanical ventilation during a case.The unit was replaced and case resumed.There was no report of patient injury.
|
|
Manufacturer Narrative
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The software was reloaded, calibrations made to resolve the issue.
|
|