H4: the lot was manufactured from july 11, 2020 - july 13, 2020.H10: the device was received for evaluation.Unaided visual inspection was performed which observed solution and leaking in a sealed bag.A functional testing was performed which revealed a leak at the spike port bonding area.The reported condition was verified.The cause of the condition could not be determined; however the most likely cause is due to inadequate or lack of cyclohexanone applied to the spike port cap tube when it was inserted to the spike port during the manufacturing process causing an incorrect bonding.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|