• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 1000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 1000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938739
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 01/25/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).Initial reporter postal code: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 1000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag leaked from an unspecified location.It was further reported that approximately 10ml was observed in the outer bag upon opening the overpouch.The bag contained synthamin parenteral nutrition solution.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from july 11, 2020 - july 13, 2020.H10: the device was received for evaluation.Unaided visual inspection was performed which observed solution and leaking in a sealed bag.A functional testing was performed which revealed a leak at the spike port bonding area.The reported condition was verified.The cause of the condition could not be determined; however the most likely cause is due to inadequate or lack of cyclohexanone applied to the spike port cap tube when it was inserted to the spike port during the manufacturing process causing an incorrect bonding.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11338985
MDR Text Key232828990
Report Number1416980-2021-00703
Device Sequence Number1
Product Code LHI
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2023
Device Catalogue NumberH938739
Device Lot Number60249585
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2021
Date Manufacturer Received03/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-