Device was evaluated.During inspection the device serial number showed (b)(4) when turned on.Further inspection found faulty pkrf board causing the current volt limit to be low and cannot be adjusted.The device hand switch lh (left-hand) and rh (right hand) failed testing due to faulty auxiliary board.Plasma blend board was observed to be faulty and unit was found running old software version and needs to be upgraded.Review of fault log showed error code 400 ref 12, one time indicated: footswitch mode pedal stuck.The usual cause is that the relevant foot pedal is held down during power on self-test or there is a faulty foot pedal, error code 400 ref 25, one time indicated: biomed time credit expired.Biomed is notified that time credit is expired.Error code stored in log.The biomed test cable has been used beyond its time limit allocation, error code 100 ref 12, one time indicated: nonvolatile memory corrupt or not initialized.Cpu checks the eeprom can be written to and read from.This error can occur immediately after a software upgrade.The identified parts were replaced, device was repaired and software was upgraded.Once completed, the device was tested and passed required testing and specifications.The investigation is ongoing; therefore, the root cause of the reported failures cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|