Model Number 50000000L
Device Problems Insufficient Cooling (1130); Failure to Pump (1502)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the arctic sun device came down from the floor saying it was not working. When they attempted a calibration, it failed for low flow with 3216 ml/min and stated would ship a replacement loaner. Per evaluation, it was reported that the arctic sun device was not cooling down and had to prime to fill which was confirmed in decontamination. The root cause was determined to be mixing pump and need service, then the arctic sun device filled with no issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11343776
MDR Text Key232389706
Report Number1018233-2021-00614
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000L
Device Catalogue Number50000000L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 02/18/2021 Patient Sequence Number: 1
-
-