• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL FMS VUE PUMP-SHAVER BOX; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL FMS VUE PUMP-SHAVER BOX; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Model Number 284004
Device Problems Display or Visual Feedback Problem (1184); Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the affiliate in (b)(6) that during an arthroscopy procedure on an unknown date, it was observed that the suction functionality and the display on the fms vue pump-shaver box were not working.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the device was received and evaluated at the service center.Neither the fault reported by the customer that the display and the suction doesn't work, could be verified nor another fault was found with the device upon evaluation.The device was cleaned, tested and found to be fully functional.Since the reported condition is not confirmed, the root cause for the reported failure cannot be determined.No manufacturing record evaluation is required as no manufacturer or service related issues were found with the device.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP-SHAVER BOX
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
MDR Report Key11344913
MDR Text Key232551372
Report Number1221934-2021-00581
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705027934
UDI-Public10886705027934
Combination Product (y/n)N
PMA/PMN Number
K171237
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284004
Device Catalogue Number284004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2021
Date Manufacturer Received02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-