Model Number E162
Device Problem Premature Discharge of Battery (1057)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  Injury  
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Device replacement was recommended but device remains in service.No adverse patient effects were reported.
 
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Device replacement was recommended.Device was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
The returned device was analyzed and a review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device battery.Low voltage capacitors are used in the high voltage charging operation in order to facilitate fast charge times.The behavior of these capacitors resulted in a high current drain, which was depleting the device battery faster than normal.This particular device was not included in the advisory population.
 
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Device replacement was recommended.Device was explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA ICD
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11355385
MDR Text Key232874780
Report Number2124215-2021-02825
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526480775
UDI-Public00802526480775
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960040/S235
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/01/2015
Device Model NumberE162
Device Catalogue NumberE162
Device Lot Number107865
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received02/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2013
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age76 YR
Patient SexMale
-
-