Model Number HST III SEAL (4.5MM), 5-PACK |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Trackwise id # (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, hst iii seal (4.5mm), 5-pack seal could not be loaded.The procedure was completed without any problems using a spare heart string.The hospital did not report any patient effects.
|
|
Manufacturer Narrative
|
Internal complaint number: (b)(4).The lot # 25154215.History record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure, hst iii seal (4.5mm), 5-pack seal could not be loaded.The procedure was completed without any problems using a spare heart string.The hospital did not report any patient effects.
|
|
Search Alerts/Recalls
|