MEDOS INTERNATIONAL SARL HD EPSCP,4.0,30,167,MITEK; RIGID ENDOSCOPE Back to Search Results
Model Number 242018
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2020
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Investigation summary: the device was received and evaluated at the service center.The reported complaint that the scope was cracked, was confirmed.The following defects were found with the device upon evaluation : the distal end of the endoscope is damaged.The whole endoscope shows usual traces of usage (i.E scratches).The distal end shows traces of usage (i.E.Discolorations, deposits).Deep scratches and dents can be seen at the distal end.The device was repaired, tested and found to be working according to specifications.The damages at the distal end are caused by applying too much force on the endoscope during usage time by the customer or by an handling error.Most probably, the distal end of the endoscope got in contact with sharp instruments during usage time or the endoscope was hit or fell down by mistake.There are no indications from this complaint investigation that the failures are manufacturing-related, therefore a manufacturing record evaluation is not required.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the sales rep that during a shoulder repair procedure on (b)(6) 2020, it was observed that the hd epscp,4.0,30,167, mitek scope device was cracked.During in-house engineering evaluation, it was determined that the distal end on the device showed traces of usage i.E.Discolorations and deposits.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD EPSCP,4.0,30,167,MITEK
Type of Device
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key11358798
MDR Text Key232860666
Report Number1221934-2021-00604
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705029150
UDI-Public10886705029150
Combination Product (y/n)N
PMA/PMN Number
K080560
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number242018
Device Catalogue Number242018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2021
Date Manufacturer Received02/22/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-