Model Number LTV 1200
Device Problem Failure to Cycle (1142)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The equipment was received in vyaire facility for evaluation and the service technician was not able to verify customer's reported problem "will not cycle".The ventilator passed 72 hours extended tests at customer settings without any unusual alarms or conditions.However, the ventilator failed troubleshooting final test at flow & o2 (oxygen) blending.As a resolution, the o2 blender was replaced and the issue was resolved.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
It is reported to vyaire medical that the ltv - lap top ventilator 1200 experienced failure to cycle.At this time, there is no information regarding patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LTV 1200 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, IL 60045
8333273284
MDR Report Key11360024
MDR Text Key232901540
Report Number2021710-2021-13423
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00845873000364
UDI-Public(01)00845873000364(11)20061229
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101643
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTV 1200
Device Catalogue Number18888-001-99
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/29/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-