Catalog Number RF320J
Device Problems Malposition of Device (2616); Patient Device Interaction Problem (4001)
Patient Problem Abdominal Pain (1685)
Event Date 05/03/2018
Event Type  Injury  
Manufacturer Narrative
Manufacturing review: the device history record review could not be performed as the lot number is unknown. Investigation summary: the device was not returned for evaluation. Medical records were provided and reviewed. Approximately eight years post filter deployment, computed tomography revealed the tip of the filter was 1. 8 centimeters below the level of the lowest renal vein. There was 6 degrees of tilt of the inferior vena cava filter. The apex of the filter was not touched the inferior vena cava walls. Filter struts were seen perforated through the wall of the inferior vena cava but did not enter adjacent structures. One of the struts abutted the right iliac artery but did not perforate the vessels. Therefore, the investigation is confirmed for perforation of the inferior vena cava (ivc). However, the investigation is unconfirmed for filter tilt as the medical record states ¿there was 6 degrees of tilt of the inferior vena cava filter and the apex of the filter was not touched the inferior vena cava walls. ¿based upon the available information, the definitive root cause is unknown. Labeling review: a review of product labeling documents (e. G. Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism. At some time post filter deployment, it was alleged that the filter tilted and struts perforated into organs. The device has not been removed and there were no reported attempts made to retrieve the filter. The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameG2 FILTER SYSTEM - JUGULAR
Type of DeviceVENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key11363892
MDR Text Key233036568
Report Number2020394-2021-80183
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberRF320J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/23/2021 Patient Sequence Number: 1
Treatment
COUMADIN
-
-