Model Number 799-AEV1-00-01 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed "internal comm failure - 1173" and "internal comm failure - 1474" messages.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the customer's report was observed and attributed to a system interconnect flex cable that was not properly seated.The cable was reseated to correct the report.The device was recertified and returned to the customer analysis for reports of this type has not identified an increase in trend.
|
|