Model Number ACU0T0
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2020
Event Type  malfunction  
Manufacturer Narrative
Complaint history records were reviewed.Product history records were reviewed and the documentation indicated the product met release criteria.The root cause has not been identified.There have been no other complaints in the reported lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported, during intraocular lens (iol) insertion with a preloaded iol the plunger went over the iol.The procedure was completed the same day with a back up iol and without patient impact.
 
Manufacturer Narrative
The device with the lens was returned.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger has advanced almost to the fill line and has overrode the lens.The lens was located within the lens loading area.The trailing haptic was folded onto the optic along the left side of the plunger.The leading haptic was ahead of the optic, located underneath the plunger.A qualified viscoelastic was indicated.The root cause cannot be determined for the reported event.The used device was evaluated.A plunger override was observed.The lens was still within the lens loading area.The plunger had been advanced almost to the fill line.A qualified viscoelastic was indicated the manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key11368427
MDR Text Key234496387
Report Number1119421-2021-00399
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/05/2023
Device Model NumberACU0T0
Device Lot Number15060552
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2021
Date Manufacturer Received03/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNSPECIFIED PROVISC OVD, 000451, UNK, BEL
-
-