Related manufacturer report numbers: 3005334138-2021-00089, 2184149-2021-00044 during an a-flutter ablation procedure, the amplifier would blink yellow when coming on ablation.The amplifier error message "logout and restart amplifier and system" was noted.The dws, generator and amplifier were all power cycled but the issue remained.The catheter was replaced and cryo was used to continue the procedure.Upon further inspection, bent pins were noted on the navx connector on the amplifier.In addition, the ensite dws secondary monitor output (via display port) was not working.The secondary image was unable to come up on siemens monitor.The display port dongle was replaced twice, in addition to replacing the dvi to dvi cable, but that did not resolve the issue.The set up was connected on the primary monitor port and the image was able to be displayed.The procedure was delayed due to these issues.There were no adverse patient consequences.
|