DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0576-01
Device Problems Disconnection (1171); No Apparent Adverse Event (3189)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console generated an iab disconnected alarm while the tubing was still connected.The customer switched out the console, but received the same message while they were attempting to move the patient in bed.Therapy was able to be continued using the second console.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
Device available for eval changed from yes to no.Product will not be returning.The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).H3 : device not returned.
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console generated an iab disconnected alarm while the tubing was still connected.The customer switched out the console, but received the same message while they were attempting to move the patient in bed.Therapy was able to be continued using the second console.There was no patient harm or adverse event reported.
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console generated an iab disconnected alarm while the tubing was still connected.The customer switched out the console, but received the same message while they were attempting to move the patient in bed.Therapy was able to be continued using the second console.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 8FR. 50CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key11372416
MDR Text Key234075145
Report Number2248146-2021-00132
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108605
UDI-Public10607567108605
Combination Product (y/n)N
PMA/PMN Number
K112327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 06/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/23/2023
Device Model Number0684-00-0576-01
Device Catalogue Number0684-00-0575
Device Lot Number3000128022
Was Device Available for Evaluation? No
Date Manufacturer Received06/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CS300 / SI226219C3CS300 / SA06416B7
Patient Age77 YR
Patient Weight64
-
-