MAQUET CRITICAL CARE AB SERVO-I; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number SERVO-I
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 02/09/2021
Event Type  malfunction  
Event Description
It was reported that the ventilator failed flow transducer test during pre-use check.There was no patient involvement.Manufacturer's ref #: (b)(4).
 
Manufacturer Narrative
The ventilator was investigated by our field service engineer.The moisture trap on the expiratory cassette was found damaged and was replaced.The valve membrane on the expiratory cassette was cleaned.The ventilator then passed all functional and safety tests and was cleared for clinical use.The replaced part was not returned for investigation.Provided ventilator logs confirms the reported failed flow transducer test during pre-use check.The root cause of the event was the damaged moisture trap and debris on the expiratory cassette valve membrane.This will be detected during pre-use check.
 
Event Description
Manufacturer's ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-I
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key11377852
MDR Text Key233466271
Report Number8010042-2021-00494
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K123149
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSERVO-I
Device Catalogue Number6487800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-