MAQUET CRITICAL CARE AB SERVO-AIR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number SERVO-AIR
Device Problem Output Problem (3005)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that the ventilator failed the pressure transducer test during pre-use check.There was no patient involvement.Manufacturer' s ref.#: (b)(4).
 
Manufacturer Narrative
Our field service technician investigated the ventilator device on site.It was reported that the ventilator failed the pre-use check during the pressure transducer test.The fault was isolated to the turbine module which was replaced and returned for investigation.After the turbine module was replaced, the ventilator passed pre-use check and was cleared for clinical use.Simulated use testing of the received turbine module has reproduced the described customer issue.An evaluation of the received device logs could also confirm the event.Our investigation concluded that a defect turbine in the turbine module caused the unit to fail the pre-use check.A technical error indicating timing issues with the turbine control was detected in the device logs.A similar error was reproduced during the ventilation tests.Replacement of the turbine resulted in a successful system pre-use check.A defective turbine module may lead to a stop of air supply during ventilation.When this error occur, the device will automatically switch to 100% o2 supply.If no o2 is available, the issue will lead to stop of ventilation.The conclusion in this matter is that the reported issue was caused by a faulty turbine in the turbine module.The root cause of the turbine failure has not been determined by this investigation.
 
Event Description
Manufacturer ref # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-AIR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
MDR Report Key11377856
MDR Text Key233466379
Report Number8010042-2021-00491
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K192604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSERVO-AIR
Device Catalogue Number6882000
Was Device Available for Evaluation? No
Date Manufacturer Received05/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-