H10: one actual device was received for evaluation.The other sample was not received and therefore, could not be evaluated.Visual inspection was performed using the naked eye which observed that the male luer was missing.Additional inspection observed a lack of solvent applied on the tubing.Dimensional testing was performed on the tubing and no issues observed.The reported condition was verified.The cause of the condition could not be determined; however, the potential causes were noted to be human error during the manual assembly step or improper performance from the solvent dispenser during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|