• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8537
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2021
Event Type  malfunction  
Manufacturer Narrative
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the connecting end (luer) of two (2) clearlink system continu-flo solution sets were missing.This was identified during priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: one actual device was received for evaluation.The other sample was not received and therefore, could not be evaluated.Visual inspection was performed using the naked eye which observed that the male luer was missing.Additional inspection observed a lack of solvent applied on the tubing.Dimensional testing was performed on the tubing and no issues observed.The reported condition was verified.The cause of the condition could not be determined; however, the potential causes were noted to be human error during the manual assembly step or improper performance from the solvent dispenser during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11378664
MDR Text Key233749666
Report Number1416980-2021-00869
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048970
UDI-Public(01)00085412048970
Combination Product (y/n)Y
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8537
Device Lot NumberR20J06128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2021
Date Manufacturer Received04/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-