• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW HEMOPRO 2
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/16/2021
Event Type  malfunction  
Manufacturer Narrative
Trackwise id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2, as the c-ring was being retracted into the cannula the left side of the c-ring stayed out and bent 90 degrees.They finished the case, did not pull another kit.The hospital did not report any patient effects.
 
Manufacturer Narrative
Trackwise # (b)(4).The lot # 25153426 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.The device was not returned to maquet cardiac surgery for investigation, however a photograph was provided by the account.A photographic inspection was conducted on 04/07/2021.Signs of clinical use and evidence of blood was observed on the intact c-ring with the one end of the c-ring appearing to be melted away from the base of the c-ring.The plastic tube was observed to be bent at an angle with the tip of the tube showing some melting.Based on the returned condition of the device, the reported failure ¿material twisted/bent; c-ring¿ as well as the analyzed failure "break, c-ring¿ were confirmed.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2, as the c-ring was being retracted into the cannula the left side of the c-ring stayed out and bent 90 degrees.They finished the case, did not pull another kit.The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
MDR Report Key11378841
MDR Text Key233556051
Report Number2242352-2021-00181
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/04/2022
Device Model NumberVASOVIEW HEMOPRO 2
Device Catalogue NumberC-VH-4000
Device Lot Number25153426
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-