Model Number N/A |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp and was able to reproduce the reported issue.The fse noticed that the battery in slot 2 would not eject and replaced the conical springs.Pm was performed during the same service visit.Full calibration, functional, and safety checks were performed which passed to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during an internal iabp training by customer, the cardiosave rescue would not eject the battery in slot 2.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during an internal iabp training by customer, the cardiosave rescue would not eject the battery in slot 2.There was no patient involvement, and no adverse event reported.
|
|