• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE RESCUE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE RESCUE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Difficult to Remove (1528)
Patient Problem No Patient Involvement (2645)
Event Date 02/11/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp and was able to reproduce the reported issue.The fse noticed that the battery in slot 2 would not eject and replaced the conical springs.Pm was performed during the same service visit.Full calibration, functional, and safety checks were performed which passed to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during an internal iabp training by customer, the cardiosave rescue would not eject the battery in slot 2.There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that during an internal iabp training by customer, the cardiosave rescue would not eject the battery in slot 2.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE RESCUE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11380560
MDR Text Key234220570
Report Number2249723-2021-00414
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108407
UDI-Public10607567108407
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-83
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-