AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 18FR DILATOR; DH EF PERC PLACEMENT PRODUCTS
|
Back to Search Results |
|
Model Number 98431 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The sample is reported to be available, but has not yet been received by the manufacturer.The device history record for lot 20016171 was reviewed and the product was produced according to product specifications.All information reasonably known as of 25 feb 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).Device not returned.
|
|
Event Description
|
It was reported that the doctor was performing a gastrostomy tube insertion procedure "last thursday" and one of the sutures in the kit snapped very easily.No further information was provided.
|
|
Manufacturer Narrative
|
The product involved in the report has been returned and the investigation remains in progress at this time.All information reasonably known as of 22 mar 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
Four sutures from the kit were returned for evaluation.All four were visual inspected and all four were broken and separated from their components.The complaint is confirmed as reported, however, no root cause could not be determined.All information reasonably known as of 13 apr 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Search Alerts/Recalls
|
|
|