Concomitant medical products: product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2019, product type: lead.Product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2019, product type: lead.Product id: 977a260, serial/lot #: (b)(4), ubd: 02-aug-2023, udi#: (b)(4).Product id: 977a260, serial/lot #: (b)(4), ubd: 26-jul-2023, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.Caller told by pt that they are charging more often.Pt does have a short circuit on 2/10 but pt was not programmed on these together coming into session today.A revision of entire system is being planned with dr.(b)(6) doing revision.Pt had a fall about a month ago and another fall about 2 weeks ago and since then pt's stimulation has felt different.While checking impedances today, #8 shows as an open circuit and 2-10 are a short at 180 ohms.Coming into the session today, pt was programmed with 8 and 2.Caller was able to reprogram pt today and get stim on pt's left side from pt's back to pt's feet (target areas) but on right side, the stimulation wouldn't go past pt's right thigh.Xray shows that lead has migrated from bottom of t7 to top of t9.They are planning a revision and planning to replace all components.
|