• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97715
Device Problems Migration or Expulsion of Device (1395); Unintended Collision (1429); Energy Output Problem (1431); Low impedance (2285); Charging Problem (2892)
Patient Problems Fall (1848); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2021
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2019, product type: lead.Product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2019, product type: lead.Product id: 977a260, serial/lot #: (b)(4), ubd: 02-aug-2023, udi#: (b)(4).Product id: 977a260, serial/lot #: (b)(4), ubd: 26-jul-2023, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.Caller told by pt that they are charging more often.Pt does have a short circuit on 2/10 but pt was not programmed on these together coming into session today.A revision of entire system is being planned with dr.(b)(6) doing revision.Pt had a fall about a month ago and another fall about 2 weeks ago and since then pt's stimulation has felt different.While checking impedances today, #8 shows as an open circuit and 2-10 are a short at 180 ohms.Coming into the session today, pt was programmed with 8 and 2.Caller was able to reprogram pt today and get stim on pt's left side from pt's back to pt's feet (target areas) but on right side, the stimulation wouldn't go past pt's right thigh.Xray shows that lead has migrated from bottom of t7 to top of t9.They are planning a revision and planning to replace all components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key11380835
MDR Text Key234976467
Report Number3004209178-2021-03269
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2020
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received02/24/2021
Date Device Manufactured09/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
-
-