• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED MOBILE INFANT WARMER; FMT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED MOBILE INFANT WARMER; FMT Back to Search Results
Model Number IW910
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint iw910 mobile infant warmer was received at our fisher & paykel healthcare (f&p) service centre in (b)(4) and was visually inspected by a trained f&p technician.Our investigation is thus based on the information provided by our service centre in france.Results: during the service, the head case of the subject iw910 mobile infant warmer was found to be cracked.Conclusion: based on our knowledge of the product and this failure mode, it is likely that the observed damage on the warmer head is related to a known problem of incompatible cleaning solutions reacting with the polycarbonate plastic and acrylic components of the infant warmer.When the heater head begins to warm up during use, a chemical reaction with the incompatible cleaning solution results in gradual crazing and cracking of the heater head.It should be noted that the subject device is over twelve years old.The infant warmer service manual for the affected unit features a diagram of the infant warmer which highlights polycarbonate and acrylic components and states the following: "caution do not clean the highlighted plastic surfaces with proprietary cleaning products containing either hydroxides, hypochlorites, peroxides, gluteraldehyde or cleaning products with a greater than 30% alcohol base" "caution the chemicals used in these proprietary cleaning products may lead to discolouration, crazing and eventual cracking of the highlighted plastic surfaces." the user manual for the iw910 mobile infant warmer states "ensure all warmer parts and accessories are checked before returning the device to service".The device technical/service manual contains a checklist which specifies that users perform safety, performance and functional checks at least once a year.
 
Event Description
A healthcare facility in (b)(6) reported of an issue with the iw910 mobile infant warmer.Upon device assessment at the fisher & paykel healthcare (f&p) regional office in (b)(4) , it was discovered that the head case was cracked.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBILE INFANT WARMER
Type of Device
FMT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, auckland 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, auckland 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key11382820
MDR Text Key252165968
Report Number9611451-2021-00221
Device Sequence Number1
Product Code FMT
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K970432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIW910
Device Catalogue NumberIW910
Device Lot Number080530
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-