BARD PERIPHERAL VASCULAR, INC. G2 FILTER SYSTEM - FEMORAL VENA CAVA FILTER Back to Search Results
Catalog Number RF310F
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Abdominal Pain (1685)
Event Date 02/05/2019
Event Type  Injury  
Manufacturer Narrative
A device history record review could not be performed as the lot number is unknown. Investigation summary: the device was not returned for evaluation. Medical records were provided and reviewed. Approximately five months later, xr-thoracolumbar spine 2 views showed the filter with tip at the level of mid l2. Eventually two years and ten months post filter deployment, the patient presented with shortness of breath and chest pain. A computed tomography (ct) revealed that no intraluminal pulmonary arterial filling defects were identified on thin section images or maximum intensity projection images. On the same day, the patient presented with abdominal pain. Subsequent computed tomography (ct) revealed that there was an inferior vena cava filter just below the level of renal veins. Eventually five years later, the patient presented with abdominal pain and a computed tomography revealed that an inferior vena cava filter was in place and located 2. 5 cm below the left renal artery. The struts did extend beyond the confines of the wall of the inferior vena cava by approximately 2mm. A medial strut abutted the wall of the abdominal aorta. Therefore, the investigation is confirmed for perforation of the inferior vena cava (ivc). Based on the available information, the definitive root cause is unknown. Labeling review: a review of product labeling documents (e. G. Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient in conjunction with trauma situation/motor vehicle accident. Approximately nine years post filter deployment, it was alleged that the filter struts perforated into organs and the patient reportedly experienced abdominal pain. The device has not been removed and there were no reported attempts made to retrieve the filter. The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameG2 FILTER SYSTEM - FEMORAL
Type of DeviceVENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key11385488
MDR Text Key233776453
Report Number2020394-2021-80211
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062887
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation
Type of Report Initial
Report Date 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberRF310F
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/26/2021 Patient Sequence Number: 1
-
-