SMITH & NEPHEW, INC. SYN TRIAL NECK SZ 14-18; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71366414 |
Device Problems
Break (1069); Corroded (1131)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that, the magnets of the syn ho trial neck sz 14-18 are corroded, chipped, and falling out.Incident occurred during service; therefore, no patient was involved.
|
|
Manufacturer Narrative
|
Results of investigation: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be chipped and corroded, rendering the device inoperable.The device was manufactured in 2014 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.Internal reference number: case-2021-00039850-3.
|
|
Search Alerts/Recalls
|
|
|