• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ASR ACETABULAR CUPS 46; HIP ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ASR ACETABULAR CUPS 46; HIP ACETABULAR CUP Back to Search Results
Catalog Number 999800746
Device Problems Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Pain (1994); Loss of Range of Motion (2032); Osteolysis (2377); Inadequate Osseointegration (2646); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/08/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Litigation records received.Ligation alleges personal and economic injuries, elevated cobalt and chromium levels, adverse local tissue reaction, pain, loss of mobility, illnesses, pseudotumor, and injuries.No labs were provided for the alleged elevated metal ions.Doi: (b)(6) 2009, dor:(b)(6) 2019, right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : previous investigations that have included device history reviews since the asr platform was launched have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.Therefore, if lot codes are provided, no dhr (device history record) review or complaint database search for this individual asr component will be carried out at this time.E3 initial reporter occupation: lawyer.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.E3 initial reporter occupation: lawyer.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Operative notes stated that there was a large effusion, there was granulomatous inflammation involving the hip capsule.The abductors were alive and viable.There was mild osteolysis around the proximal medial femur.Revision notes stated that the capsule was patulous and expanded.There was corrosion both at the trunnion and adapter sleeve.The acetabular component was easily removed without bone loss.We inspected the outer surface of the acetabular component.There was no bone on growth, only fibrous material present.The previous acetabular component was in poor position and placed with a high hip center.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : previous investigations that have included device history reviews since the asr platform was launched have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.Therefore, if lot codes are provided, no dhr (device history record) review or complaint database search for this individual asr component will be carried out at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR ACETABULAR CUPS 46
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key11386231
MDR Text Key233791679
Report Number1818910-2021-03902
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2013
Device Catalogue Number999800746
Device Lot Number2650306
Was Device Available for Evaluation? No
Date Manufacturer Received09/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ADAPTER SLEEVE 11/13 +6; ASR UNI FEMORAL IMPL SIZE 41; S-ROM*SLEEVE PRX ZTT, 18B-SML; SROM*STM ST,30+4L NK,18X13X160; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL HEAD METAL ASR; UNK HIP FEMORAL SLEEVE ASR; UNKNOWN HIP FEMORAL STEM; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL HEAD METAL ASR; UNK HIP FEMORAL SLEEVE ASR; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age67 YR
-
-