Model Number 1054260 |
Device Problems
Contamination (1120); Failure to Cycle (1142)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/30/2021 |
Event Type
malfunction
|
Event Description
|
The manufacturer received information alleging a ventilator would not cycle properly.There was no harm or injury reported.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.
|
|
Manufacturer Narrative
|
The manufacturer previously reported receiving information alleging a ventilator would not cycle properly.There was no harm or injury reported.The device was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device's blower motor, sensor board, active exhalation control module, flow sensor, and tubing need replaced to address the issue.There was evidence of contamination.
|
|