Model Number 101/860/085CZ |
Device Problems
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Event Description
|
Information received a smiths medical tracheostomy|pvc - portex tubes bluselect malfunctioned.The customer confirmed that the product worked properly in a pre-use check.However, during the use of the product, the customer noticed the inflation line got detached from the pilot balloon.No patient injury.
|
|
Manufacturer Narrative
|
Other, other text: one used decontaminated sample 10009163-005 8.5mm deht blu suctionaid sub-assy was received in plastic bag without its original packaging (see photo of sample).Under visual inspection we noticed that pilot balloon was detached from inflation line.10001429-005 infl.Line s/assy soft-seal 8.5mm is assembled in smiths medical tijuana therefore secondary investigation object in-0099123 was created and complaint sample was resent to this site.
|
|
Event Description
|
Investigation completed and summary.
|
|
Search Alerts/Recalls
|