TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25REC
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2021
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.
 
Event Description
The user facility reported to terumo cardiovascular that during a non-clinical activity, they noticed a damaged / smashed stop cock.No patient involvement as this was found during an inspection.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 26, 2021.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h4 (device manufacture date) h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307) type of investigation code#1: 10 - testing of actual/suspected device type of investigation code #2: 11 - testing of device from same lot/batch retained by manufacturer type of investigation code #3: 3331 - analysis of production records investigation findings: 3259 - improper physical structure investigation conclusions: 4307 - cause traced to component failure the affected sample was inspected upon receipt to show damage to the reservoir lid of the unit.A representative retention sample was reviewed to show no damage to the reservoir, manifold or oxygenator.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key11387597
MDR Text Key249309368
Report Number1124841-2021-00028
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450820
UDI-Public(01)00699753450820
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model Number3CX*FX25REC
Device Catalogue NumberN/A
Device Lot NumberYL21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Was the Report Sent to FDA? No
Date Manufacturer Received03/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-