MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION POLYFLUX 140H; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION POLYFLUX 140H; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 103530
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2021
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with a polyflux 140h, an external dialysate leak occurred and an ungluing was observed at the header.It was reported that there was massive dripping of dialysate and "uncontrolled dehydration".A "dialysate circuit leaking" was triggered by the control unit.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.The visual inspection by naked eye of product showed the product wet.A leak test of the dialysate side was performed and a leak was observed.Two cracks in the welding zone were also observed.The reported condition was verified.The cause of the crack could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 140H
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11387898
MDR Text Key250897845
Report Number9611369-2021-00030
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
K043342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number103530
Device Lot Number0-5719-H-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2021
Date Manufacturer Received03/01/2021
Is This a Reprocessed and Reused Single-Use Device? No