Model Number 8300
Device Problems Break (1069); Contamination (1120); Crack (1135); Failure to Cycle (1142); Communication or Transmission Problem (2896); Appropriate Term/Code Not Available (3191)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Although requested, the affected device has not been received.A follow up report will be submitted with investigation results should the device be received for evaluation.
 
Event Description
It was reported by the customer that the device received alarm - error codes / messages, error code 570.5600.There was no patient involvement.
 
Event Description
It was reported by the customer that the device received alarm - error codes / messages, error code 570.5600.There was no patient involvement.
 
Manufacturer Narrative
This device was evaluated and repaired through the service repair process.Upon visual inspection the service technician noted that they replaced oriidon pcb for error 570.6200.Replaced broken front case.Replaced dented rear case.Replaced etco2 door for won't close/stuck.Replaced contaminated left iui.Replaced broken right iui.The failure code othe was used to track the alaris pump software version as received from the customer and the software version when device was sent back to the customer.It does not reflect a device failure or represent any risk to the patient.A review of the device history record showed the device had a manufacture date of 11sep2014.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record in sap for sn (b)(6) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.Based on the findings, service determined that the probable root cause of the reported issue was due to electrical failure of the oridion board assembly.A review of the complaint history record in trackwise and sap was performed for the sn (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS ETCO2
Type of Device
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key11387933
MDR Text Key233897651
Report Number2016493-2021-29150
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10885403830013
UDI-Public10885403830013
Combination Product (y/n)N
PMA/PMN Number
K031741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8300
Device Catalogue Number8300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2021
Date Manufacturer Received06/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2722-2020
Patient Sequence Number1
-
-