The certas valve was received for evaluation.Unique device identifier (udi) (b)(4).Review of the history device records for the product code 82-8800pl with lot 4837631 conformed to the specifications when released to stock failure analysis - the valve was visually inspected; the silicone housing was cut/torn in the needle chamber.The position of the cam when valve was received was at setting 4.The valve was hydrated.The valve was leak tested and leaked from the cut/tear in the needle chamber.The valve passed the test for programming, occlusion, refux and pressure.The root cause for the problem reported by the customer is due to the cut/tear in the silicone housing in the needle chamber this was probably caused by a sharp or pointed object coming into contact with the silicone housing, as noted in the ifu silicone has a low cut/tear resistance.
|
A facility reported a tear in a shunt reservoir.A certas plus inline valve was implanted in a patient via a ventricular peritoneal shunt on (b)(6) 2020.The valve was implanted and was being tested for function and patency, before closing, when there was no flow distally and a tear was noted in the reservoir by the surgeon and a leak was discovered.There surgical case was prolonged by an estimated 30%, and the patient received an additional 20-25 minutes of anesthesia while the surgeon swapped out the damaged valve with a new valve.The patient¿s outcome was not compromised since the issue was discovered prior to closing.A medwatch report form (b)(4) was received stating that the certas valve is broken and a hole was noted by the surgeon during surgery.A new implant was placed after the issue noted.
|