SMITH & NEPHEW, INC. DRILL GUIDE HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674001
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/04/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event Description
It was reported that, during surgery, while using outside the patient, was noticed that the drill guide handle was out of alignment.No surgical delays nor injuries to the patient were reported.It is unknown how the procedure was finished.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
Results of investigation: the device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the device appears to be excessively worn, rendering it inoperable.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.The device is manufactured in 2008.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL GUIDE HANDLE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11390947
MDR Text Key233947204
Report Number1020279-2021-01638
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560469
UDI-Public03596010560469
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674001
Device Catalogue Number71674001
Device Lot Number08BCT0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2021
Date Manufacturer Received03/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-