Device stored outside of the indicated conditions.Upon receipt, the device issued ¿memory full¿, prompts with a memory capacity displayed at 12 % within saverevo.The memory capacity display in saverevo represents the total ecg/icg memory used by the device and therefore corresponds to the total recording time of the device since installation.Information from the device memory showed 91 log entries which include 57 occasions in which the device was stored below the indicated minimum and 34 manual power cycles, mainly under 1-minute duration.A low of -5.3 °c was recorded on the (b)(6) 2019.
|