HEARTSINE TECHNOLOGIES LTD PACKAGE, 350P,1X PAD-PAK-03, DUTCH, 350-STR-NL-10; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number SAM 350P |
Device Problems
Unintended Power Up (1162); Self-Activation or Keying (1557); Data Problem (3196)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/05/2021 |
Event Type
malfunction
|
Event Description
|
Device switching on automatically, memory full.No patient involvement.
|
|
Event Description
|
Device switching on automatically, memory full.No patient involvement.
|
|
Manufacturer Narrative
|
Membrane tail corrosion due to storage outside of the indicated conditions.Corrosion on the membrane tail between tracks 13 (on_button) and 14 (key3_button), which had created a partial short between these lines and multiple 10-minute timeouts in the history log, resulting in memory full prompts from the (b)(6) 2020, as per the reported fault.
|
|
Search Alerts/Recalls
|
|
|