• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC ESPRIT VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC ESPRIT VENTILATOR Back to Search Results
Model Number V1000
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2021.Date of report: 01mar2021.
 
Event Description
It was reported to philips that the device had an internal oxygen failure along with some voltage being off.The device was reported as being in use at the time of the event on a patient.The initial reporter confirmed that there was no adverse patient impact.
 
Manufacturer Narrative
The customer consulted with the philips remote service engineer (rse).The customer complaint of a vent inop 9003 state was confirmed.The rse advised the customer to replace the sensor printed circuit board assembly (pcba).The customer replaced the sensor pcba to resolve the issue.The device passed testing and was returned to service.The sensor pcba was returned to philips failure investigation (fi) for testing.Visual inspection of the sensor board revealed no evidence of damage or contamination.The sensor board was tested, and no failures were identified.The high internal o2 alarm could not be duplicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESPRIT VENTILATOR
Type of Device
VENTILATOR
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RICA
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
yobana sanchez
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key11396762
MDR Text Key240452248
Report Number2031642-2021-00779
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K981072
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV1000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-