Model Number N/A |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference reports 3003853072-2021-00015 to 3003853072-2021-00021.
|
|
Event Description
|
It was reported that seven closure top threads stripped during an operation.Other plugs were used to complete the operation.There was no patient impact.This is report five of seven for this event.
|
|
Manufacturer Narrative
|
Device evaluation: visual inspection revealed that six of the seven returned devices have damaged threads, one of these also has drive mechanism damage.One of the seven has only drive mechanism damage.Potential cause: root cause was unable to be determined.This event could possibly be attributed to cross threading during use and off-axis forces applied during insertion.Dhr review per dhr review, the part was likely conforming when it left zimmer biomet control.Device use this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
|
|
Event Description
|
It was reported that seven closure top threads stripped during an operation.Other plugs were used to complete the operation.There was no patient impact.This is report five of seven for this event.
|
|