SMITH & NEPHEW, INC. TIBIAL SPACER BLOCK STANDARD; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 74012645
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2021
Event Type  malfunction  
Event Description
It was reported that, after a tka procedure was noticed the white plastic piece from inside the tibial spacer block standard missing.Incident occurred after the procedure; therefore, there was no patient involvement.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection confirms the spacer inner white ring is missing.The device show moderate signs of wear/usage.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint, however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL SPACER BLOCK STANDARD
Type of Device
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11403051
MDR Text Key234392003
Report Number1020279-2021-01765
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556003343
UDI-Public00885556003343
Combination Product (y/n)N
PMA/PMN Number
K111711
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74012645
Device Catalogue Number74012645
Device Lot Number16LM12126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Date Manufacturer Received08/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-