MAUDE Adverse Event Report: BIOSENSE WEBSTER INC CARTO 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC CARTO 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number FG540000
Device Problems Communication or Transmission Problem (2896); Patient Device Interaction Problem (4001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2021
Event Type  Injury  
Manufacturer Narrative
The hardware investigation has begun but it has not been completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrioventricular reentrant tachycardia avrt/wolff-parkinson-white (wpw)ablation.An error 1: no communication with the patient interface unit (piu) detected, was present on the carto 3 system 20 minutes into the procedure.This resulted in the cancellation of the case after the patient was under anesthesia for approximately 2 hours.The error 1: no communication with the piu detected, was present on the carto 3 system 20 minutes into the procedure.To troubleshoot the piu and workstation were rebooted, the fiber optic cable was reseated while the system was off, without resolution.The fiber optic cable was exchanged, and system rebooted again, and it was operational for 20 more minutes then the error 1 appeared again.The original fiber optic cable was then connected again, and system rebooted, error 1 persisted.During the reboot of the piu they waited 3 minutes before booting the workstation back up.The team was advised to reconnect within 15 seconds of the piu and starting a new study.The system then initialized fully without the error.Then they were advised to use the second (ethernet port 1 on the right) on the piu and rebooting.This was done, the initial study continued.The same issue recurred, and the case was aborted as a result.Since the event is life-threatening and required intervention and prolonged hospitalization to prevent permanent impairment of a body function or permanent damage to a body structure, is to be considered serious and mdr-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO 3 SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
33 technology drive
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER (ISRAEL) LTD.
4 hatnufah street
yokneam 20667 17
IS   2066717
Manufacturer Contact
gabriel alfageme
31 technology dr
irvine, CA 92618
9497898687
MDR Report Key11408872
MDR Text Key245007900
Report Number2029046-2021-00276
Device Sequence Number1
Product Code DQK
UDI-Device Identifier10846835000870
UDI-Public10846835000870
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberFG540000
Device Catalogue NumberFG540000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-