Model Number 16-02-82 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/02/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.The heater-cooler 16-02-82 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message associated to a discrepancy between temperature measures of the safety and control temperature sensors during procedure.The affected circuit, patient or cardioplegia, is still unknown.A malfunction of the patient circuit cannot be excluded.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: no service activity has been performed yet on the device.Based on results of similar investigation, the most likely root causes of the reported event are defective temperature sensors or defective processor board.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the affected device has been replaced with another one.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|