Model Number 319.091
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2021
Event Type  malfunction  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a synthes employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2021, during an unknown procedure the drill sleeve top hat broke, retrograde/antegrade femoral nail (rafn) entry reamer is dull, drive adaptor with qc for 5.0mm schanz screws and drive adaptor with qc for 6.0 schanz screws were stripped.The percutaneous depth gauge, depth gauge for locking screws, and depth gauge for cortex screws were bent.It is unknown how the procedure was completed.There was a surgical delay of fifteen (15) minutes.This report involves one (1) depth gauge for 3.5mm cortex screws.This is report 7 of 8 for (b)(4).This product complaint, (b)(4), is related to (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 3.5MM CORTEX SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key11413411
MDR Text Key239477202
Report Number2939274-2021-01168
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982190000
UDI-Public(01)10886982190000
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number319.091
Device Catalogue Number319.091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-