• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. ALLEVYN LIFE M 12.9X12.9 CTN10; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. ALLEVYN LIFE M 12.9X12.9 CTN10; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66801068
Device Problem Nonstandard Device (1420)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2021
Event Type  malfunction  
Event Description
It was reported that, during dressing change, silicone is coming away from the dressing onto the liner while removing the liner for application and also silicone is staying on the skin/wound while removing the dressing to replace with a new dressing.It is unknown when did the event happened how was the procedure finished and if there was a delay.No other complications where reported.
 
Manufacturer Narrative
We have now completed our investigation into the reported complaint.The device used in treatment has not been returned for evaluation, with no additional information provided.We have not been able to establish a relationship between the device and the event reported or determine a root cause on this occasion.It was reported that silicone is coming away from the dressing.Factors that can contribute to the reported event include, raw material issues or storage environment issues.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history file contains further instances of the reported events.An ongoing internal action is being carried out into the reported failure mode to reduce further instances of the reported event, therefore no further actions are deemed necessary into this specific complaint.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN LIFE M 12.9X12.9 CTN10
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11429785
MDR Text Key238333608
Report Number8043484-2021-00504
Device Sequence Number1
Product Code OMP
UDI-Device Identifier05000223478018
UDI-Public05000223478018
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2023
Device Model Number66801068
Device Catalogue Number66801068
Device Lot Number202016
Was Device Available for Evaluation? No
Date Manufacturer Received04/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-