|
Model Number BELLAVISTA 1000 US |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
A vyaire field service representative (fsr) went onsite and evaluated the ventilator.Fsr verified ventilator was working properly.Calibrated o2 sensor, performed circuit test, tested ventilation, and verified alarms are working properly.No root cause has been determined at this time since the investigation is still ongoing.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported that bellavista 1000 us was tested and placed on intubated patient.The device was set in volume assist control mode with 18 bpm set rate.The respiratory therapist noticed that the bpm reading was slowly drifting down to 1 but the patient received breaths from the ventilator.They swapped the ventilator into another bellavista 1000 and bpm readings worked correctly.The customer confirmed that there was no patient harm reported from this even.
|
|
Manufacturer Narrative
|
Device evaluation: g3, g6, h2, h6 and h10.Result of investigation: no device problem detected.Machine tested by vyaire field service engineer.Error could not reproducible.Verified ventilator was working properly.Calibrated o2 sensor, performed circuit test, tested ventilation, and verified alarms are working properly.
|
|
Search Alerts/Recalls
|
|
|