• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PVP MEDIUM 6.4CM X 6.4CM; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PVP MEDIUM 6.4CM X 6.4CM; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number PVPM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Diarrhea (1811); Nausea (1970); Pain (1994)
Event Type  Injury  
Manufacturer Narrative
(b)(4).To date, the device has not been returned.If the product is returned for evaluation, any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Event Description
It was reported by an attorney that the patient underwent hernia repair surgery on (b)(6) 2014 and mesh was implanted.It was reported that the patient underwent removal surgery on (b)(6) 2014 during which the surgeon noted the mesh appeared to have protruded through the hernia defect and the fascia around the tissue had also increased the size from the previous hernia.The mesh seemed loosely approximated and therefore was removed with sharp dissection.It was reported that the patient experienced severe pain, nausea, diarrhea, chills, inflammation, loss of appetite and extreme weight loss.No additional information was provided.
 
Manufacturer Narrative
Date sent to the fda: 06/08/2021.A review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Manufacturer Narrative
Date sent to the fda: 5/25/2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PVP MEDIUM 6.4CM X 6.4CM
Type of Device
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08876
MDR Report Key11449620
MDR Text Key238910410
Report Number2210968-2021-02172
Device Sequence Number1
Product Code FTL
UDI-Device Identifier10705031132405
UDI-Public10705031132405
Combination Product (y/n)N
PMA/PMN Number
K061533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2014
Device Model NumberPVPM
Device Catalogue NumberPVPM
Device Lot NumberGP8BKRZ0
Was Device Available for Evaluation? No
Date Manufacturer Received06/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-