Model Number 97715
Device Problems Battery Problem (2885); Charging Problem (2892)
Patient Problems Burn(s) (1757); Erythema (1840)
Event Date 02/22/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 97755, serial# : (b)(4), product type: recharger. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient who was implanted with an implantable neurostimulator (ins). It was reported that¿about 2 weeks ago, the patient had to try around 8 times to connect to their implantable neurostimulator (ins) to start charging their ins. The patient said that since last thursday this happened more frequently and the patient received "no device found" and the patient was unable to charge their ins and the ins was depleted as of last night. The cord where it entered the coil on the recharger antenna (rtm) got very hot last night while charging their ins and during the call the patient verified that they have a red burn mark still on their skin from the rtm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameINTELLIS
Type of DeviceSTIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key11449884
MDR Text Key238892133
Report Number3004209178-2021-03988
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 06/23/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/14/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/21/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured03/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-