• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SPATIAL FRAME STRUT EXT SHT; PIN, FIXATION, THREADED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SPATIAL FRAME STRUT EXT SHT; PIN, FIXATION, THREADED Back to Search Results
Model Number 71070200
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event Description
It was reported that a spatial frame strut ext sht was too big and bulky for a (b)(6) patient.The doctor had to improvise to complete the procedure.Incident occurred during treatment with instruments outside the patient.No delay was reported.No other complications were reported.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, one undated photo was provided for review, it shows the device in situ, however the photo does not contribute to the root cause of the reported events.No relevant supporting clinical information has been provided to assist with a clinical investigation.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to procedural/user error.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPATIAL FRAME STRUT EXT SHT
Type of Device
PIN, FIXATION, THREADED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11465545
MDR Text Key239223399
Report Number1020279-2021-01999
Device Sequence Number1
Product Code JDW
UDI-Device Identifier03596010552204
UDI-Public03596010552204
Combination Product (y/n)N
PMA/PMN Number
K994143
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71070200
Device Catalogue Number71070200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received09/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-