C.R. BARD, INC. (BASD) -3006260740 SITE~RITE PREVUE PLUS, REBURB.; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Model Number 9770086R |
Device Problems
Image Orientation Incorrect (1305); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per biomed, the image on the screen is flipped.Turned probe over and image is still flipped.
|
|
Manufacturer Narrative
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the image on the screen is flipped.Do not use gel caps.Turned probe over and image is still flipped was unconfirmed.The prevue was tested and functioned to manufacturing specifications.Attached is a picture using a needle guide showing that the image orientation is correct.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per biomed, the image on the screen is flipped.Turned probe over and image is still flipped.
|
|