MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LAG SCREW DRILL SLEEVE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LAG SCREW DRILL SLEEVE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674523
Device Problems Connection Problem (2900); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference number: case(b)(4).
 
Event Description
It was reported that during an unspecified procedure the intertan 3.2mm guiide pin sleeve and lag screw drill sleeve were noticed to no fit together easily during surgery and intertan 3.2mm guiide pin sleeve was inserted on incorrect angle.Lag screw drill sleeve would also not pass through nail properly and surgeon had to free hand drill without the use of guide.It is suspected that the lag screw drill sleeve was bent at some point.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the affected complaint device, used in treatment, was not returned for evaluation.Therefore, a product analysis could not be performed.There is no information that would suggest the device failed to meet specifications.According to clinical/medical investigation, no relevant supporting clinical information has been provided to assist with a clinical investigation.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.A review of complaint history revealed prior complaints for the listed part.A relationship, if any, between the device and the reported incident could not be corroborated.No medical documents were received for investigation.Therefore, no medical assessment can be performed at this time.The potential probable causes for this event could include but not limited to a user or procedural error.Based on this investigation, the need for corrective action is not indicated.A review of risk management files and the instructions for use found that the probable cause failures were documented appropriately.The potential probable causes for this event could include but not limited to a patient condition, user or procedural error.A review of risk management files and the instructions for use found that the probable failures were documented appropriately.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW DRILL SLEEVE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11479246
MDR Text Key239669588
Report Number1020279-2021-02127
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00885556363249
UDI-Public00885556363249
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71674523
Device Catalogue Number71674523
Device Lot NumberUNKNOWN
Date Manufacturer Received04/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-