Model Number 27093
Device Problems Failure to Run on Battery (1466); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2021
Event Type  malfunction  
Manufacturer Narrative
The device was returned to resmed for an extensive engineering investigation.The investigation methods, results, and conclusion are not finalized at this stage.When more information is available a supplemental report will be submitted.Resmed reference #: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a power loss alarm when ac power was disconnected.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Performance testing confirmed the reported complaint and revealed the device would not power on while using its internal battery.Review of the device data logs revealed occurrences of an error message (sf180) related to a battery charger fault.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the sf180 was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a power loss alarm when ac power was disconnected.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - FUKUDA
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key11485851
MDR Text Key243131023
Report Number3004604967-2021-00390
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270934
UDI-Public(01)00619498270934(10)1447912
Combination Product (y/n)N
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 06/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27093
Device Catalogue Number27093
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date05/25/2021
Date Manufacturer Received05/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-