• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10607
Device Problems Deflation Problem (1149); Difficult to Remove (1528); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2021
Event Type  Injury  
Event Description
It was reported that deflation failure, removal difficulties, and stent foreshortening occurred. The 90% stenosed target lesion was located in the distal and proximal right coronary artery. A 3. 5 x 38 mm synergy ii drug-eluting stent was deployed for treatment. The stent balloon was inflated at 18 atmospheres for 20 seconds. However, it was observed that the balloon became stuck and was not able to deflate fully. Subsequently, the balloon was deflated by the indeflator. During the removal, the deflated balloon was pulled on the stent and it caused foreshortening. An additional 3. 5x12mm synergy stent was implanted. The procedure was completed without any other issues and the patient was stable throughout the case. No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNERGY
Type of DeviceBIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11497064
MDR Text Key240118682
Report Number2134265-2021-02772
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 03/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number10607
Device Catalogue Number10607
Device Lot Number0026425861
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/16/2021 Patient Sequence Number: 1
Treatment
GUIDE CATHETER: JR4; GUIDEWIRE: SION BLUE
-
-