Emdr follow up: the failure mode of pre-activation was confirmed based on sample received for analysis showing orange stains on the original unopened package (lidding) and foam tip.Ampoules are made from onion tubing skin borosilicate type i glass, which are design to break when pressure is applied to activate applicator at a relatively low break force.When pressure is applied to the wings by a pinching force with the fingers, this activates the applicator, breaking the glass ampoule and releasing the chloraprep solution onto the foam tip.No non-conformance was noted during the manufacturing of this lot 0289306.The most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of glass, it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excessive handing.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|