CAREFUSION, INC CHLORAPREP ONE STEP HI LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL Back to Search Results
Catalog Number 930815
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2021
Event Type  malfunction  
Manufacturer Narrative
Pr (b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
 
Event Description
Product was leaking.
 
Event Description
1 leakage of products, and 2 empty packaging.
 
Manufacturer Narrative
Emdr follow up: the failure mode of pre-activation was confirmed based on sample received for analysis showing orange stains on the original unopened package (lidding) and foam tip.Ampoules are made from onion tubing skin borosilicate type i glass, which are design to break when pressure is applied to activate applicator at a relatively low break force.When pressure is applied to the wings by a pinching force with the fingers, this activates the applicator, breaking the glass ampoule and releasing the chloraprep solution onto the foam tip.No non-conformance was noted during the manufacturing of this lot 0289306.The most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of glass, it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excessive handing.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE STEP HI LITE ORANGE
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION, INC
75 n. fairview drive
vernon hills IL 60061
MDR Report Key11499844
MDR Text Key242188692
Report Number1423507-2021-00015
Device Sequence Number1
Product Code LRJ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number930815
Device Lot Number0289306
Date Manufacturer Received07/12/2021
Patient Sequence Number1
-
-